The correct and precise formulation of the intended purpose of medical devices and in vitro diagnostic medical devices (IVD) is crucial for their successful development and approval. However, even the definitions of terms and the connection between intended purpose and intended use make it difficult to achieve clarity and precision.
This article reveals how to formulate an intended purpose to avoid difficulties during audits and approvals.
1. Relevance of the intended purpose
With the intended purpose, the manufacturer expresses the (medical) purpose for which he wants his device to be used. This definition is the prerequisite for many activities during development and approval:
- Qualification of the device, i.e., the decision as to whether it is a medical device or an IVD
- Classification of the medical device/IVD in accordance with the MDR or IVDR
- Deciding which standards must be taken into account during product development
- Specification of the device and its performance parameters
- Delimitation of normal and abnormal use (derivation of the use specifications)
- Product-specific risk management (risk management plan including derivation of the risk-acceptance matrix and risk analysis)
- Performing the clinical evaluation or performance evaluation, which must demonstrate that the purpose and medical benefit have been met
- Writing the post-market surveillance plan, which defines how achievement of the intended purpose will be monitored in the field
It is the task and freedom of the manufacturer to determine the intended purpose.
2. Definitions of the term “intended purpose”
Unfortunately, there is no uniform definition of the term “intended purpose.” In addition, many definitions equate “intended purpose” and “intended use.”
This chapter introduces the relevant definitions. The next chapter delimits intended purpose and intended use.
a) Definition according to the MDR or IVDR
“‘intended purpose’ means the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements and as specified by the manufacturer in the clinical evaluation;”
MDR Article 2 (12)
“‘intended purpose’ means the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements or as specified by the manufacturer in the performance evaluation;”
IVDR Article 2 (12)
In addition to the term “intended purpose,” the MDR/IVDR also uses the term “intended use.” However, neither regulation defines this term.
b) Definition in IEC 60601-1
IEC 60601-1 defines the term “intended purpose” as “use for which a product, process or service according to the specifications, instructions and information provided by the manufacturer” (IEC 60601-1:2005, 3.44).
c) Definition in ISO 14971
ISO 14971 has revised the definition of the term in its third edition (ISO 14971:2019). However, it equates intended use and intended purpose. But the annotation indicates that rather the medical purpose is meant.
“use for which a product, process or service is intended according to the specifications, instructions and information provided by the manufacturer
Note 1 to entry: The intended medical indication, patient population, part of the body or type of tissue interacted with, user profile, use environment, and operating principle are typical elements of the intended use. “
ISO 14971:2019 3.6
d) Definition by the FDA
The FDA defines the term “intended use” in 21 CFR part 801.4 as follows:
“The words intended uses […] refer to the objective intent of the persons legally responsible for the labeling of an article (or their representatives). The intent may be shown by such persons‘ expressions, the design or composition of the article, or by the circumstances surrounding the distribution of the article. This objective intent may, for example, be shown by labeling claims, advertising matter, or oral or written statements by such persons or their representatives. Objective intent may be shown, for example, by circumstances in which the article is, with the knowledge of such persons or their representatives, offered or used for a purpose for which it is neither labeled nor advertised; provided, however, that a firm would not be regarded as intending an unapproved new use for a device approved, cleared, granted marketing authorization, or exempted from premarket notification based solely on that firm´s knowledge that such device was being prescribed or used by health care providers for such use […]”
21 CFR part 801.4)
The Medical Device University includes a template and checklist to help you formulate your intended purpose and check for completeness.
3. Intended purpose and intended use
The term “intended purpose” in the broader sense covers several aspects in the case of medical devices and IVD medical devices:
- The actual medical purpose (“intended medical purpose” or intended purpose in the narrower sense), i.e., what disease or injury is intended to be diagnosed, treated, or monitored.
- The medical use determines the medical context, i.e., which users in which context of use are to use the device for which patients.
The (additional) intended use, which determines what can be done with the device as a whole or in addition to it. This includes storing, transporting, updating, or cleaning the device.

The MDR and IVDR refer to “normal use” instead of “intended use.” Both terms are to be understood synonymously.
4. Regulatory requirements
There are several regulations that you should consider in this context:
- The MDR and the IVDR require the intended purpose to be specified in the technical documentation (see chapter 5).
- ISO 13485 requires in 7.3.7: “Design and development validation shall […] ensure that the resulting product is capable of meeting the requirements for the specified application or intended use.” This presupposes that these are known.
- ISO 14971, the harmonized standard for risk management, requires the intended purpose for further risk analysis in chapter 5.2.
- IEC 62366-1 requires the creation of a “use specification” in chapter 5.1, the summary of which is referred to by some authorities as the intended purpose.
- IEC 60601-1 requires a precise specification of the application environment, which includes physical parameters such as temperature, altitude/air pressure, brightness, supply voltage(s), and the degree of contamination.
5. Contents of an intended purpose
5.1 Intended purpose of a medical device
Based on MDR Annex I, Chapter III, 23.4 b) and Annex II 1.1 a), address the following aspects in your intended purpose:
- Medical purpose: What disease or injury is to be diagnosed, treated, monitored, alleviated, or predicted?
- Medical indication and contraindication
- Intended patient group(s)
- Intended body part
- Intended users
- Intended use environment
- Physical environment, e.g., brightness, noise, contamination
- Social environment, e.g., stress, shift work
- Technical environment, e.g., tools, software
- Clinical environment, e.g., sterile, wear of protective equipment
- Functionality
5.2 Intended purpose of an IVD medical device
The contents of the intended purpose of an IVD are clearly specified in Annex I, Chapter III, 20.4.1 c) and Annex II 1.1 c) of the IVDR. The intended purpose of an IVD medical device includes:
- What is detected and/or measured;
- Its function (e.g., screening, monitoring, diagnosis or diagnostic aid, prognosis, prediction, companion diagnostic);
- Note: The definitions in the guidance document GHTF/SG5/N8:2012 are helpful in determining the function of an IVD medical device.
- Specific information to be provided in the following contexts:
- Physiological or pathological state;
- Congenital physical or mental impairments;
- Predisposition to a medical condition or a disease;
- Determination of the safety and compatibility with potential recipients;
- Prediction of treatment response or reactions;
- Definition or monitoring of therapeutic measures;
- Whether it is automatic or not;
- Whether it is qualitative, semi-quantitative, or quantitative;
- The type of specimen(s) required;
- The testing population, if applicable
- The intended user and;
- in the case of companion diagnostics, the International Non-Proprietary Name (INN) of the associated medicinal product.
Manufacturers of medical devices and in-vitro diagnostic medical devices usually also subsume the intended use in the intended purpose document.
6. Avoid pitfalls
a) Inconsistent formulation of the intended purpose
The intended purpose of a device is the central starting point for many subsequent processes, as the list in chapter 1 shows. During reviews of technical documentation, we repeatedly find that the intended purpose is formulated inconsistently in the various documents. As a result, not all the necessary evidence may be provided in the clinical or performance evaluation, and the notified bodies may not accept the technical file.
We recommend to set up your technical documentation based on data rather than documents to avoid redundancies and inconsistencies. Our article, 7 tips for the successful digital transformation of medical device manufacturers, provides an initial insight.
Please also note that information on the intended purpose must be consistent in the technical documentation, the associated marketing material, and the website.
b) Claims that have not been proven
Manufacturers of medical devices and in-vitro diagnostic medical devices use the intended purpose to claim the (medical) benefit of their devices.
The MDR defines clinical performance in Article 2 (52) as “the ability of a device, resulting from any direct or indirect medical effects which stem from its technical or functional characteristics, including diagnostic characteristics, to achieve its intended purpose as claimed by the manufacturer, thereby leading to a clinical benefit for patients, when used as intended by the manufacturer;” The IVDR defines the “performance of a device” as “the ability of a device to achieve its intended purpose as claimed by the manufacturer. It consists of the analytical and, where applicable, the clinical performance supporting that intended purpose;” (IVDR Article 2 (39)). “Clinical performance,” as part of the performance of a device, is defined as “the ability of a device to yield results that are correlated with a particular clinical condition or a physiological or pathological process or state in accordance with the target population and intended user;” (IVDR Article 2 (41)).
This means that manufacturers must demonstrate the clinical benefit of their device, which they state in the intended purpose, through the clinical evaluation or performance evaluation. Manufacturers should also consider the defined patient population, the intended users, and the intended use environment. Thus, the intended purpose has a direct influence on the scope of the clinical evaluation or performance evaluation and the associated effort.
For example, manufacturers could initially limit the intended purpose to a specific patient population to reduce the effort involved in the clinical or performance evaluation. After establishing the device on the market, the manufacturer can expand the intended purpose with the underlying clinical evidence.
7. Conclusion
The intended purpose is the basis for developing your medical device or IVD. If it is missing, incorrect, incomplete, inconsistent, or incomprehensible, you are likely to encounter more than just regulatory problems during development. Typically, auditors and inspectors read this document at the very beginning. So, take the time to create it carefully!
Of course, we will be happy to support you in creating your device-specific intended purpose.
Change history:
- 2025-10-06: Revision of chapter 3 regarding the connection of the terms “intended purpose” and “intended use”
- 2025-04-10: Small editorial changes
- 2024-08-12: Fundamental revision of the article; mainly integration of IVD medical device-relevant information
- 2021-03: Article completely revised; especially the introduction and chapter 3 (definitions) supplemented and updated



This article provides a very clear and detailed explanation of the importance of defining the intended purpose and intended use for medical devices and IVDs. I appreciate how it highlights the regulatory implications and the need for consistency across technical documentation, clinical evaluations, and marketing materials. The tips on avoiding common pitfalls are especially helpful for manufacturers aiming to ensure compliance, improve patient safety, and streamline the approval process.
Thank you very much for your positive feedback. We are delighted that the article was helpful to you.
I think there is a misunderstanding in interpretation of Use and Purpose. Purpose according to Merriam-Websters dictionary is ”the reason something is done or used”. Therefore, Fig. 1 in this article could be wrong and the use of a device should be a specific intention (subgroup) of a purpose. Unless there is somewhere in relevant ISOs or Regulations defined otherwise. I look forward to your response.
Thank you very much for your valuable comment. Your observation is very understandable: from a general language perspective, purpose can indeed be seen as the reason why something is done or used, so use may appear to be a subset of purpose.
In the medical device regulatory context, however, the terminology is unfortunately not used consistently. As noted in the article:
– MDR Article 2(12) defines “intended purpose” as “the use for which a device is intended …”
– IVDR Article 2(12) uses the same approach
– ISO 14971:2019 combines the terms as “intended use / intended purpose”
– IEC 60601-1 also defines intended purpose via use
For this reason, the article explicitly points out that there is no harmonized definition across the relevant regulations and standards.
The intention of Figure 1 is therefore not to establish a strict linguistic hierarchy, but rather to provide a practical structure for the different elements manufacturers typically need to define in regulatory documentation, for example the medical purpose itself, the context of use, and other use-related aspects such as handling, storage, or cleaning.
Thank you again for raising this important point.
Dear Dr. Havlicek,
Many thanks for this interesting article.
Looking at the below definitions in the EN 62366-1:2015&AC:2015&A1:2020, I think additional confusion(?) could arise (at least to me).
Par.3 Terms and Definitions
3.9 “NORMAL USE operation, including routine inspection and adjustments by any USER, and stand-by, according to the instructions for use or in accordance with generally accepted practice for those MEDICAL DEVICES provided without instructions for use.”
Note 1 to entry: NORMAL USE should not be confused with INTENDED USE. While both include the concept of use as intended by the MANUFACTURER, INTENDED USE focuses on the medical purpose while NORMAL USE incorporates not only the medical purpose, but maintenance, transport, etc. as well.
From what I understand: Normal Use (clinical + non clinical use) ≠ Intended Use = Intended Purpose (clinical use only)
In the Annex A:
Definition 3.9 – NORMAL USE
“NORMAL USE is differentiated from CORRECT USE because a USE ERROR can occur while attempting to use a MEDICAL DEVICE in accordance with its instructions for use. NORMAL USE encompasses all foreseeable USER actions when a USER is operating a MEDICAL DEVICE according to the MANUFACTURER’S INTENDED USE for that MEDICAL DEVICE. It specifically excludes ABNORMAL USE. See also Figure B.1. NORMAL USE is simply what is expected from a USER under normal conditions of use, which includes actions that are either correct or in error, but not contrary to the design intentions of the MANUFACTURER.”
From what I understand: Normal Use (clinical + non clinical use) = Intended Use
Like in your article.
Do you agree?
Thank you.
Best regards,
Marco
Dear Marco,
Thank you very much for your thoughtful comment.
I can well understand your confusion. As noted in the article, the terms “intended purpose” and “intended use” are unfortunately not used consistently across standards and regulatory texts.
In this respect, your conclusion is correct, and I agree with your interpretation. In the context of the article, “normal use” was used in the broader sense that you describe and, as you suspected, corresponds to “intended use” in that context. It therefore includes not only the medical purpose, but also other foreseeable user actions such as inspection, adjustment, or transport.
Thank you again for your valuable observation.
Best regards,
Juliane