A lot of medical device manufacturers see the international authorization of their device as a potential hazard: the opportunities offered by new markets are offset by hard-to-calculate risks as well as the time and costs required for these authorizations.
The five steps presented in this article will help manufacturers to manage these risks better. And this is essential because a failed authorization is not the worst-case scenario.
1. The challenges you have to overcome during international authorizations
In a survey conducted by the Johner Institute, participants told us the most frequent difficulties they encountered when obtaining international authorization for their devices.

The authorization process often drags on as a result of simple communication difficulties. The reasons for this are complex and range from language barriers, to different understandings of requirements, through to cultural customs.
Finding concrete and correct requirements is also always a hurdle. Regardless of whether it is provided by local distribution partners involved in the authorization or by authorities, the information is often incomplete or only provided bit by bit.
In addition, terms can be defined in different ways, which can lead to misunderstandings. And everyone who has ever gone through an authorization process knows the challenges of bureaucratic structures.
Some even get the feeling that the authorities act arbitrarily if the documentation the manufacturers have gone to great effort to create to the best of their knowledge still does not meet the requirements in the end.
In the survey, a lot of manufacturers also classified their own lack of knowledge as an obstacle. They also frequently mentioned the fact that there are different requirements in different countries as a difficulty.
Conclusion: Going through the international authorization process can be expensive, time-consuming, and frustrating.
2. 5 steps to international medical device authorizations
To prevent your authorization process from failing due to these challenges, the Johner Institute recommends going through the authorization process in five steps. These steps will help you to
- plan the international authorization in a structured way,
- avoid unforeseen costs and work, and
- minimize risks.

Step 1: Project planning
Every project starts with determining objectives. This should include:
- Selecting the markets
- Selecting the devices
- Time and sequence of market entry
Identify the authorities that regulate the authorization of your medical device in the country in question. The website of the corresponding authority won’t only contain (most of) the authorization requirements but also initial information on the process and the costs.
If possible, determine the classification of your device and the corresponding authorization pathway. These aspects usually have an impact on duration and costs.
These costs include the internal costs and regulatory fees and, if applicable, the ongoing costs of maintaining the authorizations. Maintaining an authorization often costs several thousand euros per year.
If you are missing information (e.g., because it is not available in English) that you need to complete the initial project planning, use
- a local authorized representative who can communication with authorities on site,
- an authorization partner who will be responsible for the registration,
- a consultancy company that specializes in international authorizations,
- the competent authority (in particular, for devices that use novel technologies, early coordination with the authorities is beneficial in order to coordinate the authorization procedure.).
Step 2: Selecting an authorization partner
A local authorization partner is mandatory in a lot of countries. This person must, for example, carry out the registration and authorization, and communicate with the authorities as the authorized representative.
Choose your authorization partner carefully; you will often become dependent on him because the authorization is carried out in his name.
Options
Usually, a distribution partner, an independent company, or a branch in the target country are the ideal authorization partners.
In the Johner Institute survey, more than 70% of respondents stated that their distribution partner was their main point of contact for authorizations. This is followed by their own branch offices (31%). 20% of companies work with an independent company.

Advantages and disadvantages
Each type of authorization partner has advantages and disadvantages:
| Advantages | Disadvantages | |
| Distribution partner | + Possible room for maneuver when negotiating the authorization costs + Single point of contact in target country + Distribution partner already knows the device | – The authorization is (often) in the name of the distributor. You are thus placing yourself in their dependence. – The representative (in this case, your distribution partner) can name distributors in some countries. Your distribution partner might not do this. – A change of distribution partner often requires a new registration or a change of registration – Knowledge of regulatory requirements may be limited |
| Independent company | + Being independent of the distribution partner + Cooperate with experts in medical device authorizations + Can name distribution partners without changing the authorization | – Potentially higher costs – You need an additional distribution partner |
| Own branch | + Independent management of authorizations + No additional fees for license holders + Direct communication with the authorities | – Barely any disadvantages in terms of the authorization. Sometimes it can be difficult to find qualified staff with experience in medical device authorizations in the target country. |
Decision criteria
Before deciding which authorization partner is best for you, ask yourself the following questions:
- How much time and effort does changing the authorization partner require? Is it even possible?
- Is having several distributors for one device permitted?
- Who is allowed to appoint further distribution partners?
- Is it possible to have a device authorized through several partners?
If you already have a preferred partner, then check whether this partner has the necessary prerequisites to authorize the device. Important prerequisites are:
- Existence of a quality management system
- Expertise
- Authorizations by the authority
- Reliability (if necessary, request references)
Content of the contract
The contract with your authorization partner should
- ensure you get full access to the authorization documentation,
- ensure you get access to the certificate or a copy of the certificate,
- cover market surveillance obligations and how customer feedback is handled,
- cover initial and ongoing costs for the authorization, communication with the authorities, feedback, etc.
Your authorization partner is a supplier you have to evaluate according to ISO 13485:2016.
Step 3: Defining the regulatory framework
Now, determine your detailed requirements as a manufacturer and the specific device. The specifications are often available as laws, guidance documents, referenced standards, and information on official websites.
If necessary, get help from your authorization partner, the authorities, or a consultancy company.
Country-specific information can also be found in our other articles and in our overview of regulatory authorities (German).
Documentation requirements
Above all, find out
- which documents you have to submit to the authorities,
- which contents are required for these documents,
- what form and language the submission should be made in.
Please note that not all countries accept English. For user and service manuals, in particular, countries generally insist on translations into the local language.
Use qualified and certified translation service providers. Translate critical elements back into English, if necessary. It is often the case that submission documents have not been correctly translated.
Technical requirements for international authorization
It is not uncommon for compliance with country-specific versions of standards to be required. If you know these requirements in advance, you can combine the testing of these requirements, for example, for IEC 60601-1, with your regular tests.
Some countries also require additional testing for certain device groups. More detailed research with regard to the following points is often worthwhile:
- Requirements for radio modules
- Requirements for electronic devices
- Novel technologies (e.g., “recognition of clinical studies“)
Step 4: Creating documentation
Complete technical documentation is required for a quick and smooth authorization. Although almost every country has its own requirements, there are some common elements, such as
- proof of your quality management system,
- device description, including variants,
- a clearly defined intended purpose,
- results of the risk analysis,
- instructions for use and labeling,
- results of device testing, including software verification and validation.
Use the internationally accepted document structures, such as STED, which the ASEAN member states are looking for with the Common Submission Technical Documentation (CSTD).
An alternative is the ToC (Table of Contents) Format from the IMDRF, which is what Canada and China are looking for. The ToC Format also covers large parts of the technical documentation required by the MDR.
This gives you a base you can use to extract the necessary information more easily.
You or your authorization partner submit this documentation to the corresponding authority. If the submission is successful. you will receive confirmation, e.g., in the form of a certificate.
You should first check that the certificate has been issued correctly and that the manufacturer and device information is correct. If an expiry date is given, clarify the format used for this date. It is not uncommon for the month to be given first before the day.
Step 5: Maintaining the authorization
The motto “after the authorization is before the authorization” applies. Therefore, you should determine how much time you need to plan for re-registration.
In addition, a lot of countries expect the authorization to be updated if any changes are made to the device. Keep a record of when you have to send a change notification to the authorities.
In most countries, post-market surveillance is an additional requirement for maintaining an authorization. ISO 13485:2016 also requires you to surveil the requirements in countries where you are selling the device.
Determine the requirements for the reporting of incidents and recalls and establish – together with your local partners – corresponding systems.
We are your reliable authorized representative for medical devices in Europe and Switzerland, your UK Responsible Person in the United Kingdom, and your US Agent in the USA, thus ensuring your successful market access.
3. Conclusion
The international authorization of medical devices poses countless challenges. But good planning helps to identify and overcome these challenges.
A lot of markets are worth the trouble of international authorization. The five steps help control the risks and effort required for these authorizations.
What are your experiences or the biggest challenges you had to master? We want to hear your experiences of entering new markets, e.g., either in the comment field or by e-mail.
Do you have any questions on international authorization procedures? The Johner Institute’s team will be happy to support you in every step and, if required, will also take care of the complete authorization process for you.