The EU plans to replace the existing Directive 2006/66/EC with the new EU Regulation on batteries and waste batteries (“Battery Regulation”).
This article clarifies,
- which medical device manufacturers and which medical devices are affected by the Battery Regulation,
- what the requirements of this regulation are, and
- how manufacturers can best proceed to meet these requirements.
1. Background
a) Objective
With the Battery Regulation, the EU aims to modernize the legal framework for batteries. The regulation is a response to calls for a comprehensive revision and expansion of EU legislation covering the entire life cycle of batteries.
The objective is to increase resource efficiency and promote circular value creation. The Battery Regulation is an integral part of the European Green Deal.
b) Current status of the Battery Regulation
Regulation (EU) 2023/1542 of the European Parliament and of the Council of 12 July 2023 on batteries and waste batteries, amending Directive 2008/98/EC and Regulation (EU) 2019/1020 and repealing Directive 2006/66/EC was published by the EU in the Official Journal on July 28, 2023.
Download EU Battery Regulation, first reading status June 14, 2023, English
2. Scope of the Battery Regulation
a) Which devices are affected?
Article 1 of the Battery Regulation specifies its scope of application.
(1) This Regulation lays down requirements on sustainability, safety, labelling, marking and information to allow the placing on the market or putting into service of batteries within the Union. It also lays down minimum requirements for extended producer responsibility, the collection and treatment of waste batteries and for reporting.
(3) This Regulation applies to all categories of batteries, namely portable batteries, starting, lighting and ignition batteries (SLI batteries), light means of transport batteries (LMT batteries), electric vehicle batteries and industrial batteries, regardless of their shape, volume, weight, design, material composition, chemistry, use or purpose. It shall also apply to batteries that are incorporated into or added to products or that are specifically designed to be incorporated into or added to products.
In this context, the Battery Regulation provides the following definitions.
a battery that is sealed, weighs 5 kg or less.
The regulation defines:
means any electrical or electronic equipment, as defined in Article 3(1), point (a), of Directive 2012/19/EU, which is fully or partly powered by a battery or is capable of being so;
According to Annex I, Directive 2012/19/EU (Waste Electrical and Electronic Equipment) includes all medical devices or accessories, except for all implanted and infectious products.
The Battery Regulation thus affects all medical devices (implantable and infectious devices excepted) in which device batteries are installed.
There is a single exception for medical devices in the requirement in Article 11 for the removability and replaceability of device batteries.
Examples of medical devices that are affected:
- Syringe pump with a backup battery
- Transportable patient monitor
- Removable laryngoscope handle with battery
- Mobile therapy robot with an electric drive powered by a transaction battery
- Ear thermometer with replaceable button cell battery
- Blood pressure monitor for home use with standard AAA cells
Examples of medical devices that are not affected:
- Implantable cardiac pacemakers with rechargeable battery (specifications are included in the MDR)
- Surgical instruments that must be disposed of via hospital waste
- Disposable laryngoscopes with illumination and fixed battery that must be disposed of via hospital waste
Standalone software
For software medical devices installed and operated on general mobile platforms or tablets, the manufacturer of the hardware is responsible for compliance with the regulation. It may be necessary for the manufacturer of the software medical device who delivers the software together with a hardware device to include battery labeling information in the medical device user manual.
b) Which organizations are affected?
First, the Battery Regulation affects all manufacturers who produce or develop a battery (including batteries installed in appliances or vehicles) or have them produced and market that battery under their own name or trademark or put it into service for their own purposes and market it for trade or use for the first time in the territory of a member state.
Manufacturers also count as producers, importers, or distributors!
That the term “manufacturer” is broadly defined is shown by Article 3, Paragraph 47, which states that the following activities qualify an organization as a “manufacturer”:
- Established in the EU: producing or designing or having produced and distributed in the EU under its own name or trademark
- Established in the EU: reselling batteries produced by other EU manufacturers if the name or trademark of these other manufacturers is not indicated
- With establishment in the EU: first commercial sale of batteries originating from another member state or a third country
- With establishment in the EU or a third country: direct sale via distance contracts to end users
Which medical device manufacturers are affected?
Medical device manufacturers fall under the first condition if they
- have a battery developed and produced for their medical device (e.g., a patient transport monitor) to meet their needs and market it with the device under their own name.
- develop and produce a finished usable battery from battery cells for the medical device themselves.
Which medical device manufacturers are not affected?
Manufacturers whose devices are designed for standard batteries and where no batteries are supplied are not affected. However, the instructions for use must contain information on the type of battery to be used. This is already a requirement of the MDR.
3. The main requirements of the regulation
The Battery Regulation comprises 13 chapters and 14 annexes (see Fig. 1).

The requirements for medical device manufacturers are:
- Carbon footprint statement
- Removability and replaceability of device batteries
- Labeling and information requirements
- Obligations of economic operators
a) Statement on the carbon footprint
Article 7 requires manufacturers for electric vehicle batteries, rechargeable industrial batteries with a capacity greater than 2 kWh, and LMT batteries to submit the carbon footprint in the form of a declaration in the future. The methods for calculating the CO2 footprint will be specified by the EU.
The introduction of the regulations over time will take place in three stages:
- Stage 1 – Obligation to provide information on the CO2 footprint in the form of a declaration
- Stage 2 – Classification of batteries into performance classes
- Stage 3 – Introduction of maximum values for the CO2 footprint over the entire life cycle
Initially, only manufacturers of certain types of batteries will be affected by the requirements:
- Rechargeable industrial batteries with internal storage and a capacity greater than 2 kWh
- Electric vehicle batteries
- LV batteries (light transportation)
The deadlines for the different battery types are regulated in Article 7, Subsections 2 and 3. The requirements do not apply to portable batteries.
Medical device manufacturers who produce batteries or have them developed or produced are also considered manufacturers. They are, therefore, subject to these regulations.
b) Removability and replaceability of device batteries
The Battery Regulation requires that the batteries can be removed and replaced:
Any natural or legal person that places on the market products incorporating portable batteries shall ensure that those batteries are readily removable and replaceable by the end-user at any time during the lifetime of the product. That obligation shall only apply to entire batteries and not to individual cells or other parts included in such batteries.
Article 11, Paragraph 1
This means that manufacturers must ensure the following characteristics of the batteries:
- Replaceable if the life of the batteries is shorter than that of the device
- Removable at the latest at the end of the service life
- End users or independent economic operators must be able to remove or replace the battery
Any natural or legal person that places on the market products incorporating portable batteries shall ensure that those products are accompanied with instructions and safety information on the use, removal and replacement of the batteries.
For the following devices, battery replacement may be limited to independent professionals:
- Appliances specifically designed to operate primarily in an environment that is regularly subject to splashing water, water streams, or water immersion, where such derogation is required to ensure the safety of the user and the appliance (e.g., sterilizable medical devices)
- Appliances where the battery can only function if it is integrated into the device structure (e.g., shared enclosure, hearing aids where applicable)
- Professional medical imaging and radiotherapy devices as defined in Article 2, No. 1 of Regulation (EU) 2017/745 and in vitro diagnostic medical devices as defined in Article 2, No. 2 of Regulation (EU) 2017/746
It is unclear why the EU includes all IVDR devices in the exemption but only professional medical imaging and radiation therapy devices in the medical devices. Perhaps it is a mistake, and all medical devices are meant?
Article 11(3) of the EU Battery Regulation provides an important exemption for certain medical devices: the requirement for portable batteries to be removable and replaceable by end-users does not apply where continuity of power supply is necessary and a permanent connection between the product and the respective battery is required – either for safety reasons for the user and the appliance or for data integrity reasons in products whose main function is to collect and supply data.
The European Commission specifies in its guidelines from January 2025 that this exemption applies particularly to life-saving, life-sustaining, and safety-critical medical devices. Explicitly mentioned are implantable medical devices such as cardiac pacemakers and implantable cardioverter defibrillators, certain high-risk class in-vitro diagnostic devices, and hearing aids where battery replacement could pose a safety risk to patients. Medical devices with CMOS batteries for maintaining volatile memory or backup functions also fall under the data integrity exemption. The Commission recommends a risk-based approach using the classification systems of the MDR (Article 51, Annex VIII) and IVDR (Article 47, Annex VIII) to determine the applicability of the exemption.
c) Label and information requirements
As of January 1, 2027 batteries must be visibly, legibly, and indelibly marked. Labeling includes the indication of certain key characteristics such as lifetime, charge capacity, obligation for separate collection, the presence of hazardous substances, and safety risks.
A permanent QR code is to be placed on batteries (printed or engraved), which must provide access to information relevant to the battery in question, the battery passport, and carbon footprint information, depending on the battery type.
Stationary battery energy storage systems, LMT batteries, and electric vehicle batteries shall include a Battery Management System (BMS) that stores the information and data necessary to determine the aging status and expected lifetime of batteries manufactured in accordance with the parameters set out in Annex VII. This requirement can already be derived from the MDR (see MDR Annex I, Paragraph 18.2).
Paragraphs 1 and 2 of the new Battery Regulation concern only producers of batteries. Paragraph 3 concerns manufacturers using battery types for which the BMS is mandatory and who must provide an interface for the readability of the information.
d) Electronic exchange of information: the battery passport
For the electronic exchange of information, manufacturers of LV batteries, industrial batteries, and electric vehicle batteries should create an electronic file (“battery passport”). These requirements only apply to manufacturers who are themselves producers of a battery. The battery passport contains information on the battery model and specific information for the individual battery in accordance with Annex XIII. The battery passport is made accessible online via a QR code.
The information to be stored in the European electronic exchange system shall include:
- Publicly available information according to Annex XIII, No. 1
- Information accessible only to notified bodies, market surveillance authorities, and the Commission in accordance with Annex XIII, No. 2 and No. 3
- Information accessible only to natural or legal persons in accordance with Annex XIII, No. 2 and No. 4
The regulation also specifies the requirements for the structuring of the information and the interoperability of the data exchange. The information is based on open standards and is presented in an interoperable format. It is machine-readable, structured and searchable. Article 78 describes the technical design of the battery passport.
The requirement only applies to manufacturers that develop and produce batteries and are considered battery manufacturers according to the regulation.
e) Obligations of the economic operators
Chapter VI of the new Battery Regulation sets out the obligations of economic operators.
| Economic operator | Article | Activities abstract |
| Manufacturers | Article 38 | Fulfill sustainability and safety requirements according to Chapter II Preparation of the technical documentation according to Annex VIII Labeling according to Article 13 Carrying out conformity assessment Always ensure conformity for series products (standards, Common Specifications) Provide access to values according to Annex VII in the battery management system Corrective action and market surveillance Cooperation with national authorities |
| Distributors | Article 42 | Checking the registration of the manufacturer Checking the CE marking of the battery Checking whether the operating instructions and safety information are enclosed Check whether the producer and the importer have fulfilled their requirements In case of non-conformity, inform the producer and the market surveillance authority |
| Importers | Article 41 | Check conformity with Articles 6 to 10 and 12, 13, and 14 before placing on the market Before placing on the market, check that the battery bears a CE marking In case of non-conformity inform the producer and the market surveillance authority |
| Authorized representatives | Article 40 | Keeping the EU declaration of conformity available Keeping available the report on the review of the due diligence strategies, confirmed by a notified body Handing over to the authority all the information and documentation required to demonstrate conformity Cooperating with national authorities on all measures to avert the risks |
| Suppliers | Article 39 | Provide producers with the information and documentation necessary to comply with the requirements Note: Manufacturers/producers must ensure that they receive this information |
4. Battery Regulation and medical device manufacturers
a) Applicability to medical device manufacturers
Manufacturers who purchase and use batteries ready for use
Manufacturers who purchase and use ready-to-use batteries must comply with the following requirements:
- Check the label
Verify that sufficient information is available on the intended use of the battery to enable it to be properly commissioned, used, and managed at the end of its life. These requirements are indirectly contained in IEC 60601-1, Section 4.8 on the use of components. - Verify compliance
Verify that the battery bears the CE marking and ensure that the distributor provides the information (operating instructions and safety information) in accordance with its obligations. Manufacturers should take appropriate measures to ensure that they only make batteries available on the market that comply with this regulation. - Providing information
Provide information to the end user on safe disposal and replacement with the necessary tools where appropriate. - Design and layout
Design equipment so that batteries (provided safety and performance are not affected) are removable and replaceable; provide means in the equipment to display or read information from the Battery Management System (BMS).
Manufacturers who develop batteries or have batteries developed
Manufacturers who produce batteries or have batteries developed or produced and who market these batteries under their own name or trademark must fulfill the following requirements:
These manufacturers (producers), when placing on the market and putting into service batteries (including for their own purposes), must ensure that
- the battery meets the sustainability and safety requirements described in Chapter II,
- the battery is labeled in accordance with Article 13,
- the technical documentation has been prepared in accordance with Annex VIII,
- a conformity assessment procedure has been carried out in accordance with Article 17, and
- the CE marking is affixed in accordance with Articles 19 and 20.
b) Implementation of the Battery Regulation
What should manufacturers do?
Manufacturers should perform the following activities:
- Check whether their activities fall within the scope of the regulation or whether the requirements for removability and replaceability of portable batteries apply to their own devices.
- Check if the obligations for distributors or importers apply to them and implement the additional obligations in the QM management system, if applicable.
- Check if the exemptions for removability and replaceability of batteries in Article 11 apply. If not, manufacturers must design the equipment and batteries so that the batteries are easily removable and replaceable.
- Provide means in the device to display or read the information from the Battery Management System (BMS).
- Implement interfaces and protocols. To our knowledge, there is no standard. There is a fact sheet from the VDE.
- Integrate the information from the battery’s producer into your own user manual or include the information.
The regulation places great emphasis on transparency and traceability to promote an environmentally friendly circular economy. Non-European manufacturers may not want to fully comply with the required requirements, especially on transparency. Therefore, manufacturers should assess long-term supply availability for batteries at an early stage. If necessary, they should consider seeking a manufacturer within the EU (which was the intention of the EU Commission).
What should manufacturers avoid?
The Commission wants to regulate the use of rechargeable and non-rechargeable general-purpose portable batteries (e.g., AAA batteries) by setting parameters for their electrochemical performance and shelf life.
The objective is to reduce the use of non-rechargeable general-purpose portable batteries. The Commission is even considering a ban.
Manufacturers should consider whether the use of a non-rechargeable general-purpose portable battery is environmentally justified over a rechargeable battery (e.g., because only limited recycling is possible).
c) Transition periods
The transition periods only affect producers of batteries:
| Carbon footprint | The deadlines depend on the battery type and are regulated in Article 7, Sections 2 and 3. Example: For electric vehicle batteries, the deadline is February 18, 2025 or 12 months after the date of entry into force either of the delegated act or of the implementing act respectively. |
| Battery labeling | General information according to Annex VI, Part A for all batteries: Article 13, Paragraph 1: From August 18, 2026 or 18 months after the date of entry into force of the implementing act referred to in Paragraph 10, whichever is the latest. Symbol referred to in Annex VI, Part B for all batteries: From August 18, 2025. QR code, Annex VI, Part C for all batteries: Article 13, Paragraph 6: From February 18, 2027. For certain types of batteries, further labels and their deadlines are regulated in Article 13, Paragraphs 3 to 10. |
| Digital battery pass | 42 months after the date of entry into force of this regulation |
5. Implementation in Germany: The Battery Implementation Act (BattDG)
a) Effective date
The BattDG has been in force since October 7, 2025. It regulates the practical implementation of the EU Battery Regulation in Germany, in particular the registration requirements and the organization of take-back and collection systems.
b) Affected medical device manufacturers
This affects all medical device manufacturers who are also considered “manufacturers” within the meaning of the EU Battery Regulation (Def. 47, BattVO). This is the case if they place products with built-in or attached batteries on the market in Germany under their own name or trademark, regardless of whether they a) developed the batteries themselves, b) had them developed on their behalf, or c) purchased them from others and resell them under their own name/brand.
c) Registration requirements
Anyone who is a manufacturer according to the Battery Regulation bears extended producer responsibility for batteries that they make available in Germany for the first time and must register with the Federal Environment Agency before placing them on the market for the first time (Section 5, BattDG). Registration takes place on the ear portal of the Stiftung Elektro-Altgeräte Register (EAR), which has been entrusted with the relevant tasks by the Federal Environment Agency.
In Germany, BattVO manufacturers must carry out the registration themselves and cannot transfer it to a producer responsibility organization (Section 7, sentence 1, BattDG).
Manufacturers based outside Germany (e.g., Switzerland) must appoint an authorized representative based in Germany (Section 40 (2) BattDG) to take care of the registration.
Manufacturers must register in each member state in which they place a battery on the market for the first time.
d) When BattVO manufacturers must register
Case 1: No registration requirement as a BattVO manufacturer exists
- if an MDR manufacturer based in Germany develops a patient monitor with an integrated backup battery and purchases the finished battery (e.g., from Varta) in Germany. The MDR manufacturer delivers the device with the built-in battery to clinics in Germany.
The MDR manufacturer is a distributor within the meaning of the BattVO and not a BattVO manufacturer, as Varta is the domestic producer of the battery and assumes the role of BattVO manufacturer. In its role as a BattVO distributor, the MDR manufacturer must check the conformity of the battery (CE marking), check the registration of the battery manufacturer, fulfill its take-back obligation for used batteries, and inform customers about correct disposal (Articles 42 and 62, BattVO).
Case 2: The registration requirement as a BattVO manufacturer applies
- if an MDR manufacturer based in Germany develops and distributes an ultrasound device with a special lithium-ion battery. The battery is manufactured in China according to the manufacturer’s specifications, bears the manufacturer’s logo, and is sold together with the device in Germany.
The MDR manufacturer is a manufacturer within the meaning of the BattVO (BattVO manufacturer): He …
- places the battery on the market in Germany for the first time under his own name.
- must register with the Federal Environment Agency before placing the product on the market for the first time.
- bears extended producer responsibility and must participate in a take-back system in Germany, report annual quantities, ensure labeling, and provide information on disposal.
e) Further links
- ear information page on registration
- Battery Implementation Act – BattDG
- Federal Environment Agency list of topics
6. Summary and conclusion
With the Battery Regulation, the EU wants to contribute to environmental protection. If your medical devices contain batteries, you should check carefully whether you are affected by the regulation and comply with the legal requirements.
Irrespective of this, you can contribute to environmental protection by, for example, refraining from using non-reusable batteries.
The Johner Institute helps manufacturers find out if they are affected by the regulation and define and implement a regulatory strategy.
Get in touch if you want to find out how we can support you to get your medical devices into all markets quickly and in compliance with the law.
Change history
- 2025-12-10: Section 5 on implementation in Germany added
- 2025-12-18: Clarification in section 3 b) when obligations regrading removability and replaceability of portable batteries shall not apply



Reading the possible derogations for Article 11, I think the section in this article could be improved. Paragraph 3 of Article 11 states ‘The obligations laid down in paragraph 1 shall not apply where continuity of power supply is necessary and a permanent connection between the product and the respective portable battery is required to ensure the safety of the user and the appliance or, for products that collect and supply data as their main function, for data integrity reasons.’ I believe this applies to many portable/wearable medical devices and could be referenced clearly here to help manufacturers.
Dear Mr. Woodcraft,
Thank you for this important comment. I have reworked the wording of the section and added some considerations from the European Commission that could be taken into account when assessing the applicability of Article 11.
best regards, Mario Klessascheck
Am I right to assume that if we integrated a simple button-battery on our main board which is contained in almost any “stationary” wall-powered equipment we’re also in scope?
In practive nobody will open up the device to find that small on-board battery to remove it.