Medical device manufacturers must determine the requirements for their devices and prove that these are met. This often leads to regulatory problems, partly because there are misunderstandings about the different types of requirements.

Content

This page provides a quick overview of:

  1. Types of requirements
  2. Regulatory requirements
  3. Possibilities of support for elicitation and verification

There are links to further articles on all topics.

1. Types of requirements

During the various development phases, the different types of requirements must be identified and fulfilled (see Fig. 1).

The V-model shows the phases in which the requirements are collected. Stakeholder requirements (green); technical requirements (blue)

Fig. 1: The V-model shows the phases in which the requirements are collected. Stakeholder requirements (green); technical requirements (blue)

The stakeholder and technical requirements are subdivided into further types (see Fig. 2).

Fig. 2: The stakeholder requirements (green) and the technical requirements (blue) each have subtypes.

Fig. 2: The stakeholder requirements (green) and the technical requirements (blue) each have subtypes.

Further information

These articles present the different requirement types with the respective methods for elicitation and verification as well as the regulatory requirements in detail:

2. Regulatory requirements

There are regulatory requirements for the requirements, specifically for the elicitation, tracking and verification of the requirements.

a) Requirements for the collection of requirements

The regulatory requirements for medical devices oblige manufacturers to collect and provide evidence of these requirements.

Examples

  • The MDR requires the determination of the relevant essential requirements. It requires acceptance criteria, also for clinical requirements and claims.
  • IEC 62304 insists on acceptance criteria for the verification of software units.
  • IEC 62366-1 requires acceptance criteria, specifically for evaluating the user interface.
  • IEC 60601-1 specifies numerous acceptance criteria (e.g., maximum leakage currents and temperatures), also known as test criteria.
  • ISO 13485 demands that customers’ requirements be determined, both those that are explicitly mentioned and those not mentioned. It obliges manufacturers to identify all regulatory requirements.

b) Requirements for the tracing of requirements

The laws and standards also demand that manufacturers trace the requirements. Manufacturers should distinguish between two types of traceability:

  1. Horizontal traceability
    Horizontal traceability (dotted red lines in Fig. 3) is intended to ensure that compliance with all requirements is actually verified and validated.
  2. Vertical traceability
    Vertical traceability (dashed red lines in Fig. 3) is intended to ensure that the requirements are actually taken into account during further development.
Man unterscheidet horizontale (gepunktete Linien) und vertikale (gestrichelte Linien) Traceability.

Fig. 3: A distinction is made between horizontal traceability (dotted lines) and vertical traceability (dashed lines).

Traceability requires that the individual requirements are clearly identified, for example, with an ID.

c) Requirements for the proof of requirements

The laws and standards oblige manufacturers to prove that the devices meet all the requirements placed on them. Many manufacturers equate requirements with acceptance criteria. Even though this often works in practice, the two terms are not synonymous.

Acceptance criteria describe how you decide whether you consider a requirement to be fulfilled.

Example

If there is a requirement that a software unit should calculate the square of a number, then the acceptance criteria can be formulated as the test cases that are used to check this. But the test cases are not the requirement.

Tip

The Johner Institute recommends that the system requirements specify how a system behaves as a black box.

There are acceptance criteria that do not(!) affect this black box. These include, for example, compliance with coding guidelines for a software unit. This compliance would not be a requirement for the software unit itself but for its development.

The fact that the acceptance criteria are met is necessary but not sufficient to prove that the requirements are also met. In practice, however, proof of fulfilled acceptance criteria is usually enough as “objective evidence.”

Support in collecting and verifying the requirements

Do you have questions about the requirements? You will receive answers via our free micro-consulting.

Contact us if you would like support. We will happily help you quickly, conveniently, and in compliance with the law to collect and verify all relevant requirements. This will ensure that your devices are safe and in demand and can be launched on the market quickly.


Use scenario – user story – user task: finally getting the big picture

IEC 62366-1 uses the concept of (hazard-related) use scenario. The FDA recognizes critical (user) tasks. In development, use cases and user stories are used. This variety of terms makes it difficult to achieve a common understanding. However, this is necessary to create a consistent and legally compliant usability engineering file and avoid regulatory problems. This…

Details